Home / Products / Oxygen Sensors (O2) / SX-GO2EGA Series
Ir directamente a la información del producto
SX-GO2EGA Series
1/2

SX-GO2EGA Series

The SensorsX SX-GO2EGA Series is a compact galvanic electrochemical oxygen sensor designed for accurate O₂ concentration measurement. It provides a 0–25% VOL detection range, fast response, low power consumption, and stable performance, making it suitable for industrial safety, medical oxygen equipment, air-quality monitoring, and oxygen detection applications.

  • Model: SX-GO2EGA-110
  • Detection Principle: Electrochemical, two-electrode (galvanic)
  • Target Gas: Oxygen (O₂)
  • Detection Range: 0–25% VOL
  • Response Time: <20 s
  • Zero Drift: 0.6% VOL
  • Resolution: 0.1% VOL
  • Operating Temperature: −30°C to +50°C
  • Service Life: 2 years
Ordering Code
$30.00

Product Overview

The SX-GO2EGA Series is a galvanic electrochemical oxygen sensor. Oxygen is reduced at the working electrode and an oxidation reaction occurs at the counter electrode, generating an electrical current. The output current is proportional to the oxygen concentration, allowing O concentration to be determined from the measured current.

Main Features

  • High Accuracy, Long Service Life  —— Precise detection with extended operating life
  • Fast Response, Fast Recovery —— Quick response with rapid baseline recovery
  • Low Power, High Sensitivity —— Low power consumption with high sensitivity
  • Wide Linear Range —— Robust performance across the measurement range
  • High Resistance to Interference —— Low susceptibility to common interfering gases 
  • High Repeatability and Stability —— Consistent long-term performance 

Product Application

Oxygen alarms for mining, petrochemical, and related industries
Medical oxygen concentrators
Air-quality monitoring instruments
Underground parking garage and commercial air purification equipment
In-cabin air-quality monitoring and vehicle cabin airpurification / fresh-air systems

Technical Data

Measurement
Example Model SX-GO2EGA-110
Detection Principle Electrochemical, two-electrode (galvanic)
Detection Gas Oxygen (O₂)
Detection Range 0–25% VOL
Maximum Overload Concentration 30% VOL
Sensitivity & Output
Output Signal 100 ± 20 µA @ 20.9% O₂
Zero Drift 0.6% VOL
Resolution 0.1% VOL
Linearity Linear, regression coefficient R² = 0.999
Output  Current proportional to O₂ concentration
Response
Response Time < 20 s
Repeatability 2% of output signal
Long-Term Stability < 5% signal/year
Stability & Service Life
Shelf Life 12 months after delivery
Service Life 2 years
Electrical Data
Bias Voltage 0 V (self-powered galvanic cell)
Load Resistance 100 Ω
Environmental Conditions
Operating Temperature −30°C to +50°C
Operating Humidity 15–90% RH, non-condensing
Operating Pressure 90–110 kPa
Mechanical Data
Body Diameter Ø20 mm, without label
Body Height 15.9 mm
Pin Length 4.5 mm
Pin Diameter Ø1.5 mm
Dimensional Tolerance ±0.2 mm
Cross-Interference
Carbon Dioxide CO₂ 0.00% O₂ equivalent reading at 5%
Hydrogen (H₂) 0.00% O₂ equivalent reading at 2000 ppm

Certifications & Compliance

 

Related Products

Ready to upgrade your sensing infrastructure?

You can directly input your requirements and the AI ​​will automatically match the most suitable model.